10µHz–1MHz Versatile High Performance Gain/Phase Frequency Response Analyzer
PSM1700 PsimetriQ is a comprehensive DFT and true RMS analysis instrument for R&D and production test applications. Specifically aimed at a range gain/phase analysis applications, amongst which include switched mode power supply feedback loop stability analysis and general frequency response analysis of filters. It has a clear, bright graphics display showing real time values or graphs or tables. It can be operated completely by remote control over RS232 or LAN as an option. When combined with the IAI (Impedance Analysis Interface) the PSM become a high accuracy 1MHz Impedance Analyzer/ LCR meter. It is designed to give optimum performance, speed of measurement and convenience, providing sophisticated measurement capability in a cost effective and easy to use bench top instrument.
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
Control LoopStabilityAnalysis |
TransistorPerformanceAnalysis |
Filter Design |
Audio AmplifierDesign |
OptoCouplerEvaluation |
Coax CableFrequencyResponse |
LDO RegulatorEvaluation |
Signal TransformerPerformanceEvaluation |
Cross TalkTesting |
RFI/EMC FilterDesign |
Product Images
Features & Benefits
Frequency range 10µHz to 1MHz | Floating differential inputs – 100 Vpk |
Gain accuracy at < 1 kHz 0.02 dB | Phase accuracy at < 1 kHz 0.02° |
Frequency response analyzer | Impedance analyzer |
Vector Volt Meter / Phase Angle Volt Meter | RMS Volt Meter |
Loop response analysis | Harmonic analyzer |
Brochures | Documentation |
PSM1700_35 Brochure & Technical Specifications PDF, 747kb | PSM1700 User Manual v2.23 PDF, 1.075Mb |
Introduction to PSM1700 and PSM1735 Datalog Mode | PSM17xx Communications Manual v2.01 PDF, 1.096Mb |
LCR Active Head & IAI | IAI User Manual PDF, 769kb |
PSMComm2 Brochure | PSMComm2 User Manual v1_1c PDF, 2.79Mb |
PSM1700
PSM1700 Options, Software and Accessories
![]() |
Communications Interfaces: |
- Standard: RS232
- Optional: LAN(Option L), GPIB(Option G)
![]() |
Probes and Probe Accessory Options: |
- LCR1 Active Head + Kelvin Lead Set
- IAI Impedance analyser Interface + Kelvin Lead Set
- IAI Kelvin Fixture Axial and Radial Component connection for LCR measurements
- TA107 Transimpedance Amplifier
- TTF35 Telecom Test Fixture
- PSM1 Power Meter Adaptor (Enables Power Measurements with the PSM1700)
- LCR Kelvin Head Kelvin Replacement Lead Set
- HVx100 2500V Attenuating Probe
![]() |
Injection Transformer Options |
- Standard Injection Transformer 10Hz to 200kHz
- High Frequency Injection Transformer 500Hz to 35MHz
- High Voltage Injection Transformer 5Hz to 15MHz 600V CAT II Isolation
- Low Frequency Injection Module DC to 100kHz
![]() |
Current Measurement Options: |
- HF010m 10mΩ/20Arms 1MHz Bandwidth Current Shunt
- HF100m 100mΩ/6Arms 1MHz Bandwidth Current Shunt
- HF470m 470mΩ/3Arms 1MHz Bandwidth Current Shunt
- HF01A 1Ω/1Arms 1MHz Bandwidth Current Shunt
![]() |
Port Accessory Options: |
- USB to 9 Pin RS232 Serial Converter
![]() |
Carry Cases: |
- PSM Soft Carry Case – Black Nylon Padded Case with accessory space, handle and strap
![]() |
Rack Mounting Kit Options: |
- PSM17xx Rack Mounting Kit – 19″ Rack Mount Kit including brackets and fixing screws for PSM17xx
![]() |
Calibration and Documentation: |
- Standard Recalibration – PSM17xx
- Standard Recalibration – PSM17xx + IAI
- External Laboratory ISO17025 UKAS Calibration + Standard Calibration – PSM17xx
- Spare set of user manuals – Hard copy of user and communications manual
![]() |
PC Software (Free of charge): |
- PSMComm – PSM PC Interface Software, databasing, graphs, real time data display (PSMComm2 User Manual v1_1b PDF, 2.79Mb)
- PSMSoft – Labview Executable with PSM, including Nyquist Simulation
Application Notes and Related Articles
For application notes please see the app note section of our Knowledge Database
MEASUREMENT SPECIFICATIONS | |
---|---|
FREQUENCY RESPONSE ANALYSER | |
Measurement | Magnitude, Gain (CH1/CH2 or CH2/CH1), Gain (dB), Offset Gain (dB), Phase (⁰) |
Frequency Range | 10μHz to 1MHz 20mHz to 500kHz with External Source |
Gain Accuracy in dB | 0.02dB < 1kHz 0.05dB < 10kHz 0.1dB + 0.003⁰/kHz < 1MHz |
Phase Accuracy | 0.02⁰ < 10kHz 0.02⁰ + 0.003⁰/kHz < 1MHz |
Frequency Source | Generator or CH1 Input |
Measurement | Real Time DFT, no missing data |
Speed | Up to 100 readings per second |
Filter | Selectable from 0.2 seconds |
Resolution | 5 or 6 digits |
VECTOR VOLTMETER | |
Measurement | In-Phase, Quadrature, Tan Φ, Magnitude, Phase, In-Phase Ratio, rms, rms ratio, LVDT Differential, LVDT Ratiometric |
Frequency Range | 10μHz to 1MHz 20mHz to 500kHz with External Source |
Basic Accuracy (ac) | 0.05% range + 0.05% reading + 0.05mV < 1kHz Basic + 0.02%/kHz < 10kHz Basic + 0.2% + 0.002%/kHz < 1MHz |
LCR METER | |
Functions | L, C, R (ac), Q, tan delta, impedance, phase – Series or Parallel circuit |
Frequency Range | 10μHz to 1MHz |
Current Shunt | External or N4L active head or IAI (Impedance Analysis Interface) |
LCR Measurement Functions | |
Resistance (R) | 1mOhm to 500MOhm |
Reactance (X) | 1mOhm to 500MOhm |
Impedance (Z) | 1mOhm to 500MOhm |
Capacitance Value (Cp & Cs) | 1pF to 1mF |
Phase Angle | 0 to 360 deg with 0.001 degree resolution |
Q. Factor | +/- 0.00001 ~ 9999.99 |
Tan Delta (Reciprocal of Q, also known as Dissipation Factor D) | +/- 0.00001 ~ 9999.99 |
Admittance (Y) reciprocal of impedance (Siemens) | +/- 2nS to 1KS |
Conductance (G) resistive part of admittance (Siemens) | +/- 2nS to 1KS |
Susceptance (B) reactive part of admittance (Siemens) | +/- 2nS to 1KS |
Voltage Bias | +/- 10V |
Ranges – LCR Head | Inductance – 100nH to 10kH Capacitance – 10pF to 1000μF Resistance – 10mΩ to 100MΩ |
Ranges – IAI | Inductance – 10nH to 10kH Capacitance – 1pF to 1000μF Resistance – 1mΩ to 500MΩ |
Basic Accuracy | 0.1% + tolerance of selected current shunt |
Sweep Capability | All ac functions |
Channels | 2 |
Frequency Range | DC to 1MHz |
Measurement | rms, ac, dc, peak, cf, surge, dBm |
Basic Accuracy (ac) | As VVM + 0.2mV |
Accuracy (dc) | 0.1% range + 0.1% reading + 1mV |
IAI Impedance Analysis Interface | |
Frequency Range | DC to 1MHz |
Basic Accuracy | 0.1% < 1kHz 0.2% + 0.002% / kHz < 1MHz |
Internal Shunt Phase Accuracy | Low 5Ω 0.1deg + 0.01deg / kHz Normal 50Ω 0.05deg + 0.005deg / kHz High 5kΩ 0.05deg + 0.005deg / kHz Very High 500kΩ 0.1deg + 0.05deg / kHz |
Power Source | 90-264V rms 47-64Hz |
POWER METER | |
Measurements | W, VA, PF, V, A, total, fundamental and integrated, power harmonics |
Frequency Range | 20mHz to 1MHz |
Current Shunt | External or use N4L power adaptor |
Current Accuracy | As Voltage + external shunt tolerance |
Watts Accuracy | 0.15% VA range + 0.15% reading + external shunt tolerance |
HARMONIC ANALYSER | |
Scan | Single or Series |
Frequency Range | 10μHz to 1MHz |
Measurement | Harmonic, series THD or difference THD |
Max Harmonic | 50 |
SYSTEM SPECIFICATIONS | |
INPUT RANGES | |
Inputs | 2 Differential |
Connectors | Isolated BNC |
Coupling | ac or ac+dc |
Max Input | 100Vpk from earth |
Input Ranges | 100V, 30V, 10V, 3V, 1V, 300mV, 100mV, 30mV, 10mVpk |
Scaling | 1 x 10^-9 to 1 x 10^9 |
Ranging | Full auto, Up only or manual |
Input Impedance | 1M // 50pF (exc leads) |
SIGNAL GENERATOR | |
Type | Direct Digital Synthesis |
Frequency | 10μHz to 1MHz |
Waveforms | Sine, Triangle, Square, Sawtooth |
Accuracy Open Loop (with no trim) | Frequency ±0.05% ±10% < 100kHz ±20% < 1MHz |
Accuracy Closed Loop (with trim) | ±1% < 100kHz ±5% < 1MHz |
Impedance | 50Ω ±2% |
Output Voltage (Open Circuit) |
0V to ±10Vpk |
Output Resolution | 5mV |
Offset | 0V to ±10Vpk |
Offset Resolution | ±10VmV |
Clock Rate | 11.52MHz |
Connector | Grounded BNC |
DATALOG | |
Functions | Up to 4 measured functions user selectable |
Datalog Window | From 10ms with no gap between each log |
Memory | RAM or non – volatile up to 8,000 records |
HIGH SPEED DATA STREAMING | |
Rate | 1500 readings/s max |
Window | 660μs to 1s Synchronized to waveform |
Buffer | 8000 results |
GENERAL | |
Display | 320 x 240 dot LCD – white LED backlight |
Alarm | Any displayed function Hi, Lo, Inside Window or Outside Window |
Program Stores | 100, prog 1 loaded on power up |
Sweep Stores | 30, all parameters in any sweep function |
Remote Operation | Full capability, control and data |
Size | 170H x 350W x 250D mm approx |
Temperature | 5 to 35⁰C |
Weight | 4Kg approx |
Power Supply | 90-264V rms 47-63Hz 30VA max |