Impedance Analysis Interface: Impedance Analyzer
PSM series accessory for LCR test
The PSM1700 and PSM1735 can provide LCR measurements in either 'passive mode', where an external shunt is used for current detection or with optional impedance measurement accessories available from N4L.
Passive mode
In this mode, an external shunt is placed in series with the component under test, then CH1 is used to measure the voltage across the component under test while CH2 is used to measure the voltage across the shunt. From these voltages and their relative phase angle, the PSM units compute impedance and all LCR functions.
LCR Active Head and Impedance Analysis Interface
The LCR Active Head and IAI convert the PSM1700/1735 units into high performance LCR meters with true 4 wire Kelvin connections that are taken directly to the component under test without the need for external shunts. Buffering, amplification and selectable shunts provide LCR measurements over a wide frequency and impedance range.
| LCR Active Head | IAI | |
|---|---|---|
| Frequency Range | 10µHz to 5MHz | 10µHz to 35MHz |
| Measurement Ranges Inductance |
100nH to 10kH | |
| Capacitance | 10pF to 1000µF | |
| Resistance | 10mΩ to 100MΩ | |
| Absolute accuracy | 0.2% <1kHz | 0.1% <1kHz |
| 0.5% + 0.005%/kHz <5MHz | 0.2% + 0.0005%/kHz <35MHz | |
| Accuracy with reference component trim | 0.2% + 0.0002%/kHz | |
| Internal Shunts: Low |
N/A | 5Ω |
| Normal | 100Ω | 50Ω |
| High | 10kΩ | 5kΩ |
| Very High | 1MΩ | 500kΩ |
| Power source | Auxiliary Port | 90–264V rms 47–64Hz |
» Contact our sales department to discuss your requirements...
This product is designed, manufactured & supported in the UK. All measurements are traceable to UKAS, certificates of conformance and calibration are supplied as standard. Our products are CE marked.
N4L Impedance Analyzers...
Impedance Analysis Interface
LCR Active Head
Impedance Analysis Interface
connected to the PSM1735
LCR measurement display
Downloads
Brochure & Technical Specifications PDF, 130Kb